Display Color Analyzer CA-410 Probe Firmware Update
NOTE: The latest version of PC Software for Color Analyzer CA-S40 / SDK for Color Analyzer CA-SDK2 must be used to use any new functions added when the probe firmware is updated. Click here to go to the software update page.
Changes from previous version
Changes from Ver. 1.30 to Ver. 1.31
Bug fixes:
The following bug has been fixed:
- When measurement speed is set to AUTO and low-luminance measurements of a single color are taken, abnormal values may be shown.(This bug exists only in firmware ver. 1.30.)
Changes from Ver. 1.20 to Ver. 1.30
Product specifications:
- A function to automatically detect the frequency of the measurement subject in the range of 10 to 240Hz has been added.
- A function to automatically detect the frequency of the measurement subject in the range of ±2Hz from the frequency value (in the range of 10 to 240Hz) set by the user has been added.
- A function to enable the shortest integration time (Single-Frame) for synchronized measurement to be selected has been added.
- A function to automatically retry zero calibration when a zero calibration error has occurred has been added.
Changes from Ver. 1.11 to Ver. 1.20
Product specifications:
- Compatibility with Probes CA-VP402, CA-VP404 and CA-VP410T has been added.
Bug fixes:
The following bug have been fixed:
- Rare problem in which communication errors occurred the first time the probe was connected to the computer.
Changes from Ver. 1.10 to Ver. 1.11
Product specifications:
- Compatibility with Ø10 Probe (CIE 170-2:2015 model) CA-P410C and Ø27 Probe (CIE 170-2:2015 model) CA-P427C has been added.
Bug fixes:
The following bugs have been fixed:
- Various minor bugs.
Changes from Ver. 1.03 or prior versions to Ver. 1.10
Product specifications:
- Waveform measurement function has been added.
- High-frequency flicker measurement function has been added.
- Zero calibration time has been increased by approx. 0.4 s (from approx. 3.4 s for Ver. 1.03 or prior version to approx. 3.8 s for Ver. 1.10) due to addition of new functions.
NOTE: PC Software for Color Analyzer CA-S40 / SDK for Color Analyzer CA-SDK2 must be Ver. 1.30 or later to use the newly added functions. Click here to go to the software update page.
Bug fixes:
The following bugs have been fixed:
- Problem in which the error of chromaticity measurement values was slightly increased in rare cases when used in a high-temperature environment.