The essentials of imaging

  • Global
  • Measuring Instruments
search form

Non-contact 3D Digitizer

Measuring Instruments

content

VIVID 9i

Non-contact 3D Digitizer

VIVID 9i provides high precision, making it well suited for the dimensional measurement demands of industrial processes such as dies, casting, forging, sheet metal and molded plastics part production.

Specifications

Principal Specifications

Type Non-contact 3D digitizer
Measuring Method Triangulation light block method
Light-Receiving Lenses (Interchangeable) TELE Focal distance f=25 mm
MIDDLE Focal distance f=14 mm
WIDE Focal distance f=8 mm
Scan Range 0.6 to 1.0 m (In Standard mode)
0.5 to 2.5 m (In Extended mode)
Laser Scan Method Galvanometer-driven rotating mirror
Laser Class Class 2 (IEC60825-1), Class 1 (FDA)
X Direction Input Range (In Extended mode) TELE 93 to 463 mm
MIDDLE 165 to 823 mm
WIDE 299 to 1495 mm
Y Direction Input Range (In Extended mode) TELE 69 to 347 mm
MIDDLE 124 to 618 mm
WIDE 224 to 1121 mm
Z Direction Input Range (In Extended mode) TELE 26 to 680 mm
MIDDLE 42 to 1100 mm
WIDE 66 to 1750 mm
Accuracy (X,Y,Z) ±0.05 mm (Using TELE lens at distance of 0.6 m, with Field Calibration System, Konica Minolta's standard, at 20°C)
Precision (Z, σ ) 0.008 mm (Using TELE lens at distance of 0.6 m, Konica Minolta's standard, at 20°C)
Input Time (per scan) 2.5 sec
Transfer Time to Host Computer Approx. 1.5 sec
Ambient Lighting Condition Office environment, 500 lx or less
Imaging Element 3D data: 1/3-inch frame transfer CCD (340,000 pixels)
Color data: Common with 3D data (color separation by rotary filter)
Number of Output Pixels 3D data/Color data: 640 x 460
Output Format 3D data: Konica Minolta format, & (STL,DXF, OBJ, ASCII points, VRML)
(Converted to 3D data by the Polygon Editing Software/ standard accessory)
Color data: RGB 24-bit raster scan data
Data File Size Total 3D and color data capacity: 3.6MB per data
Viewfinder 5.7-inch LCD (320 × 240 pixels)
Output Interface SCSI II (DMA synchronous transfer)
Power Commercial AC power, 100 to 240 V (50/60Hz), rated current 0.6 A (at 100 VAC)
Dimensions 221 (W) × 412 (H) × 282 (D) mm
Weight Approx. 15 kg (with lens attached)
Operating temperature/humidity range 10°C to 40°C, relative humidity 65% or less with no condensation
Storage temperature/humidity range 0°C to 40°C, relative humidity 85% or less (at 35°C) with no condensation
Regulatory approvals UL 61010A-1, CSA-C22.2 No.1010-1, etc.
  • Specifications are subject to change without notice.
  • Product names in this brochure are trademarks of their respective companies.

Back to Top